Outcomes of a metal-on-metal total hip replacement system

2014 
IntroductionHigh short-term failure rates have been reported for a variety of metal-on-metal (MoM) total hip replacements (THRs) owing to adverse reactions to metal debris (ARMD). This has led to the withdrawal of certain poorly performing THRs. This study analysed the outcomes of a MoM THR system. MethodsBetween 2004 and 2010, 578 uncemented MoM THRs (511 patients, mean age: 60.0 years) were implanted at one specialist centre. The THR system used consisted of the Corail® stem, Pinnacle® cup, Ultamet® liner and Articul/eze® femoral head (all DePuy, Leeds, UK). All patients were recalled for clinical review with imaging performed as necessary. ResultsThe mean follow-up duration was 5.0 years (range: 1.0–9.1 years). Overall, 39 hips (6.7%) in 38 patients (all 36mm femoral head size) underwent revision at a mean time of 3.5 years (range: 0.01–8.3 years) from the index THR with 30 revisions (77%) performed in women. The cumulative eight-year survival rate for all THRs was 88.9% (95% confidence interval [CI]: ...