Accurate Noninvasive Estimation of Left Ventricular End-Diastolic Pressure: Comparison with Catheterization

1998 
Abstract We evaluated the accuracy of a new Doppler-based method using the mitral regurgitant velocity at the time of aortic valve opening for the noninvasive estimation of left ventricular end-diastolic pressure. Sixty unselected patients were studied immediately before routine catheterization. Invasive left ventricular end-diastolic pressure was obtained using a fluid-filled pig-tail catheter. Noninvasive estimation of left ventricular pressure at aortic valve opening was taken as systemic diastolic pressure using an automated cuff. Noninvasive left ventricular end-diastolic pressure was calculated as diastolic blood pressure - 4 × (mitral regurgitant velocity at aortic opening) 2 . Those making noninvasive determinations were blinded to catheterization results. An adequate mitral regurgitant Doppler recording was obtained in 24 patients (40%). In patients with a left ventricular end-diastolic pressure greater than 15 mm Hg the yield was 65%. Left ventricular end-diastolic pressures ranged from 4 mm Hg to 30 mm Hg. Bland and Altman analysis revealed no systematic bias and close agreement was found, with individual discrepancies not exceeding 5 mm Hg. (J Am Soc Echocardiogr 1998;11:126–31.)