Safety, effectiveness and cost efficacy of diagnostic electrophysiology and radiofrequency ablation in a district general hospital.

2001 
Electrophysiological studies (EPS) are now being performed in district general hospitals (DGH) in the UK. In order to audit our results, a prospective database was established for all patients undergoing EPS and radiofrequency (RF) ablation at Eastbourne District General Hospital, East Sussex. Between 1 January 1997 and 1 July 2000, 300 EPS procedures were performed, resulting in 155 RF ablations. The average RF ablation procedure time was 119.3 minutes with an average fluoroscopy time of 19.1 minutes. Cost per RF ablation procedure was £ 1166.79 excluding use of facilities, pacemaker devices, medical nursing and radiography staffing costs. The overall success rate for RF ablation was 93.6% with a major complication rate of 0.6%, a total complication rate of 3.9% and no associated mortality. We have shown that RF ablation can be performed safely, effectively and economically in a DGH setting with a high rate of success and a low complication rate.