Mechanisms of failure and survival of total femoral endoprosthetic replacements

2019 
Aims The aim of this study was to evaluate the prosthesis characteristics and associated conditions that may modify the survival of total femoral endoprosthetic replacements (TFEPR). Patients and Methods In all, 81 patients treated with TFEPR from 1976 to 2017 were retrospectively evaluated and failures were categorized according to the Henderson classification. There were 38 female patients (47%) and 43 male patients (53%) with a mean age at diagnosis of 43 years (12 to 86). The mean follow-up time was 10.3 years (0 to 31.7). A survival analysis was performed followed by univariate and multivariate Cox regression to identify independent implant survival factors. Results The revision-free survival of the implant was 71% at five years and 63.3% at ten years. Three prostheses reached 15 years without revision. The mean Musculoskeletal Tumor Society score in the group was 26 (23 to 28). The mechanisms of failure were infection in 18%, structural failures in 6%, tumour progression in 5%, aseptic loosening in ...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    20
    References
    18
    Citations
    NaN