Design and Optimization of a Dual-HPGe Gamma Spectrometer and Its Cosmic Veto System

2017 
In this paper, a dual high purity germanium (HPGe) gamma spectrometer detection system with an increased solid angle was developed. The detection system consists of a pair of Broad Energy Germanium (BE-5030p) detectors and an XIA LLC digital gamma finder/Pixie-4 data-acquisition system. A data file processor was developed containing five modules that parses Pixie-4 list-mode data output files and classifies detections into anticoincident/coincident events and their specific coincidence types (double/triple/quadruple) for further analysis. A novel cosmic veto system was installed in the detection system. It was designed to be easy to install around an existing system while still providing sufficient cosmic veto shielding comparable to other designs. This paper describes the coverage and efficiency of this cosmic veto and the data processing system. It has been demonstrated that the cosmic veto system can provide a mean background reduction of 66.1%, which results in a mean MDA improvement of 58.3%. The counting time to meet the required MDA for specific radionuclide can be reduced by a factor of 2-3 compared to those using a conventional HPGe system. This paper also provides an initial overview of coincidence timing distributions between an incoming event from a cosmic veto plate and HPGe detector.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    11
    References
    4
    Citations
    NaN
    KQI
    []