0460: Radiation dose reduction in pediatric coronary CT: assessment of effective dose and image quality

2015 
Aim To assess the impact of different scanning protocols on radiation dose and image quality for pediatric coronary computed tomography. Materials and methods 100 children underwent coronary computed tomography after arterial switch operation from November 2012 to March 2014. Scans were done using two different scanner models without difference in scanning and image reconstruction parameters: Light speed VCT and Discovery HD750, 64-slice from GE Healthcare. Two consecutive changes in scanning protocols were performed: 1) the use of adaptive statistical iterative reconstruction (ASIR) instead of filtered back projection (FBP) for image reconstruction; 2) the optimization of scan acquisition parameters (current and tube voltage reduction). Premedication (beta-blocker) was used for all children to obtain heart rate  Results Patients were divided in three age groups: 0–4, 5–7 and 8–18 years. After adjustment for scan settings, median ED decreased by 28% (3.9 mSv, IQR 2.8–4.2), 40% (0.9 mSv, IQR 0.6–2.6) and 65% (0.7 mSv, IQR 0.5–0.9) for 0–4, 5–7 and 8–18 years age groups ( P Conclusions Coronary CT can be obtained at very low radiation doses in pediatric patients using ASIR and prospective ECG-triggered acquisition with optimized scan parameters.