Late opacification of the foldable hydrophilic acrylic lens SC60B-OUV

2001 
Abstract Objective To investigate the cause of severe central opacification in 41 foldable acrylic intraocular lenses (IOLs) requiring explantation. Another IOL was opacified in the original sealed vial. Design Case series and laboratory analysis. Testing Light microscopy, high performance liquid chromatography, sodium dodecyl sulfate polyacoylamide gel electrophoresis, spectrometric analysis, and autoclaving. Results Neither fatty acids nor proteins could be identified within the IOLs. Spectrometric analysis yielded absorption peaks in the ultraviolet spectral range. Conclusions The spectroscopic findings indicate premature aging of the ultraviolet blocking agent. The source of the opacification is a change in the IOL material itself.