Measurement of human myocardial perfusion by double-gated flow alternating inversion recovery EPI.

1999 
This paper presents a flow-sensitive alternating inversion recovery (FAIR) method for measuring human myocardial perfusion at 1.5 T. Slice-selective/non-selective IR images were collected using a double-gated IR echoplanar imaging sequence. Myocardial perfusion was calculated after T1 fitting and extrapolation of the mean signal difference SISel - SINSel. The accuracy of the method was tested in a porcine model using graded intravenous adenosine dose challenge. Comparison with radiolabeled microsphere measurements showed a good correlation (r = 0.84; mean error = 20%, n = 6) over the range of flows tested (0.9–7 ml/g/min). Applied in humans, this method allowed for the measurement of resting myocardial flow (1.04 ± 0.37 ml/g/min, n = 11). The noise in our human measurements (SEflow = 0.2 ml/g/min) appears to come primarily from residual respiratory motion. Although the current signal-to-noise ratio limits our ability to measure small fluctuations in resting flow accurately, the results indicate that this noninvasive method has great promise for the quantitative assessment of myocardial flow reserve in humans. Magn Reson Med 41:510–519, 1999. © 1999 Wiley-Liss, Inc.