Intraoperative electromyography monitoring in minimally invasive transforaminal lumbar interbody fusion.

2007 
Object Minimally invasive transforaminal lumbar interbody fusion (TLIF) is an increasingly popular method for achieving lumbar decompression and fusion. The procedure is technically more demanding than open fusion, with correspondingly more theoretical risk of complication. The authors describe the use of intraoperative electromyography (EMG) as an adjunct to surgery to reduce the risk of complications. Methods Between August 2005 and April 2006, 25 consecutive patients underwent minimally invasive TLIF in which a total of 105 pedicle screws were placed. Intraoperative EMG was performed and included passive recordings during decompression and interbody graft placement, as well as active recording during the placement of the pedicle access needle and testing of the pedicle tap. A uniform protocol for active monitoring was used, with the pedicle access needle set at 7 mA. To assess hardware placement, all patients underwent postoperative radiography and 20 underwent postoperative computed tomography (CT) sc...