Lung Insufflation Capacity with a New Device in Amyotrophic Lateral Sclerosis: Measurement of the Lung Volume Recruitment in Respiratory Therapy

2020 
Objective The aim of this study was to validate the usefulness of the measurement of lung insufflation capacity (LIC) using the LIC TRAINER (LT) in patients with amyotrophic lateral sclerosis (ALS). Methods This retrospective study was conducted in the rehabilitation departments of the Japanese National Center of Neurology and Psychiatry and involved 20 ALS patients who underwent respiratory therapy between April 1, 2014, and December 2017. The vital capacity (VC), maximum insufflation capacity (MIC), and LIC measurements at the start of respiratory therapy were extracted from the medical records, and patients were divided into three groups: group A, VC could not be measured; group B, VC could be measured, but MIC was less than VC; and group C, MIC was larger than VC. LIC could be measured in all groups. In group C, paired t-tests were used to analyze whether there was a significant difference in the volumes measured using different methods. Results LIC was 950, 1863±595, and 2980±1176 ml in groups A (n=1), B (n=10), and C (n=9), respectively. In groups A and B, LIC could be measured in all patients, even when VC or MIC could not be measured. In group C, the measured LIC value was significantly greater than MIC (p=0.003). Conclusion LIC could be successfully measured using the LT. By using the LT, it was feasible to conveniently perform LIC measurements, suggesting that it could be a useful device for performing respiratory therapy in ALS patients.