Assessment of the MANTA closure device in transfemoral transcatheter aortic valve replacement: a single-centre observational study

2020 
The present study aims to evaluate the efficacy and safety of the MANTA vascular closure device (VCD) (Teleflex, Morrisville, NC, USA) in transfemoral transcatheter aortic valve replacement (TF-TAVR). To close the femoral artery in TF-TAVR a VCD is the treatment of choice. Data involving suture-based VCDs have been extensive. Although scarce, results on the MANTA device are promising. There is no consensus yet as to whether the MANTA device is associated with fewer access-site-related vascular/bleeding complications when compared to suture-based VCDs. In this prospective single-arm study, performed at a single centre, a total of 73 patients eligible for TF-TAVR were included and consecutively treated with the MANTA device. Access-site-related vascular complications were seen in 13.7% of patients treated with the MANTA device. In this group of patients only minor vascular complications were observed. Access-site-related bleeding complications were rare (6.8%), and device failure was seen in 13.7% of the patients. This single-centre study confirms that use of the MANTA device in TF-TAVR is feasible with an acceptable rate of access-site-related complications and no major vascular complications.