Comparison between high-frequency ultrasonography and histological assessment reveals weak correlation for measurements of scar tissue thickness

2017 
Abstract Objective Current methods for evaluating scar tissue volume following burns have shortcomings. The Vancouver Burn Scar scale is subjective, leading to a high variability in assessment. Although histological assessment via punch biopsy can discriminate between the different layers of skin, such an approach is invasive, inefficient, and detrimental to patient experience and wound healing. This study investigates the accuracy of high-frequency ultrasonography, a non-invasive alternative to histology, for measuring dermal and epidermal thickness in scar tissue. Methods Scar thicknesses of 10 patients following burns were assessed using a 2-D high-frequency ultrasound probe. The scars were then biopsied using a circular 4 mm punch biopsy for histological assessment. Dermal, epidermal, and total thickness of the scar tissue was measured using ultrasound and histology, and correlations between the two measurements were calculated. Results There was not a strong correlation between ultrasound measurement and histological analysis for epidermal, dermal, and total thickness (Spearman’s rank correlation of −0.1223, −0.6242, and −0.6242) of scar tissue. Conclusions Measurements of scar thickness using high-frequency ultrasonography did not recapitulate the in vivo dermal, epidermal and total thickness. Based on these findings, strategies for further optimization of 2-D ultrasonography is discussed before clinical and research use.