In vitro biomechanical testing of the 3.5 mm LCP in torsion: a comparison of unicortical locking to bicortical nonlocking screws placed nearest the fracture gap

2017 
This biomechanical study compared the torsional strength and stiffness of a locking compression plate with all locking versus nonlocking screws and examined the effect of placing a locking unicortical or nonlocking bicortical screw nearest the fracture gap in a synthetic bone model. Synthetic bone models simulating a diaphyseal fracture without anatomic reduction were tested using four screw configurations: all bicortical locking (ABL), all bicortical nonlocking (ABN), a hybrid construct with a bicortical nonlocking screw nearest the fracture gap (BN), and a unicortical locking screw placed nearest the fracture gap (UL). Torsional stiffness, rotation and torque at failure were compared via ANOVA and post hoc pairwise comparisons (p < 0.05). ABN and BN had the highest stiffness (p < 0.01) with ABL greater than UL (p < 0.01). Rotation at failure was greatest for ABL (p < 0.01) with UL greater than ABN (p < 0.05). Unicortical locking screws nearest the fracture gap decreased stiffness, without significantly affecting torque or rotation at failure. Construct stiffness was found to exist in a very narrow range of 0.9–1.2 N m/deg with standard deviations of 0.1 N m/deg in all cases. The results of this study support the use of nonlocking screws in a hybrid construct to increase torsional stiffness.