HPLC/PDA determination of carminic acid and 4-aminocarminic acid using relative molar sensitivities with respect to caffeine

2018 
ABSTRACTTo accurately determine carminic acid (CA) and its derivative 4-aminocarminic acid (4-ACA), a novel, high-performance liquid chromatography with photodiode array detector (HPLC/PDA) method using relative molar sensitivity (RMS) was developed. The method requires no analytical standards of CA and 4-ACA; instead it uses the RMS values with respect to caffeine (CAF), which is used as an internal standard. An off-line combination of 1H-quantitative nuclear magnetic resonance spectroscopy (1H-qNMR) and HPLC/PDA was able to precisely determine the RMSs of CA274nm/CAF274nm and 4-ACA274nm/CAF274nm. To confirm the performance of the HPLC/PDA method using RMSs, the CA and 4-ACA contents in test samples were tested using four different HPLC-PDA instruments and one HPLC-UV. The relative standard deviations of the results obtained from five chromatographs and two columns were less than 2.7% for CA274nm/CAF274nm and 1.1% for 4-ACA274nm/CAF274nm. The 1H-qNMR method was directly employed to analyse the CA and 4-A...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    16
    References
    5
    Citations
    NaN
    KQI
    []