A Comparison of the Forces Applied to a Manikin during Laryngoscopy with the Glidescope® and Macintosh Laryngoscopes:

2011 
The force applied during laryngoscopy can cause local tissue trauma and can induce cardiovascular responses and cervical spine movement in susceptible patients. Previous studies have identified numerous operator and patient factors that influence the amount of force applied during intubation. There are few studies evaluating the effect of different laryngoscope blades and no study involving video laryngoscopes. In this study we measured the forces using two laryngoscopic techniques. Three FlexiForce Sensors® (A201-25, Tekscan, Boston, MA, USA) were attached to the concave blade surface of a Macintosh and a GlideScope® laryngoscope. Experienced anaesthetists performed Macintosh and GlideScope intubations on the Laerdal® Airway Management Trainer manikin. Compared to Macintosh intubations, the GlideScope intubations had equal or superior views of the glottis with 55%, 58% and 66% lower median peak, average and impulse forces applied to the tongue base. The distal sensor registered the most force in both dev...