[The Application of Gold-immunochromatographic Test Strip Reader in Serum Antibody Detection in Echinococcosis].

2015 
One hundred and fifty-nine serum samples from hydatid disease patients and 80 serum samples from patients with other liver diseases were detected by gold-immunochromatographic assay, and read by naked eyes and the gold-immunochromatographic test strip reader. The sensitivity, specificity, and accuracy of eye-based method was 92.4% (147/159), 85.0% (68/80), and 89.9% (215/239), which was lower than that of the reader detection (95.6%, 93.7%, 95.0%, respectively). While, its false negative rate (7.5%, 12/159) and false positive rate (15.0%, 12/80) was higher than that of the reader detection (4.4% and 6.3%, respectively).