Plateletworks® as a Point-of-Care Test for ASA Non-Sensitivity.

2021 
Aspirin (ASA) therapy is proven to be effective in preventing adverse cardiovascular events; however, up to 30% of patients are non-sensitive to their prescribed ASA dosage. In this pilot study, we demonstrated, for the first time, how ASA non-sensitivity can be diagnosed using Plateletworks®, a point-of-care platelet function test. Patients prescribed 81 mg of ASA were recruited in a series of two successive phases-a discovery phase and a validation phase. In the discovery phase, a total of 60 patients were recruited to establish a cut-off point (COP) for ASA non-sensitivity using Plateletworks®. Each sample was simultaneously cross-referenced with a light transmission aggregometer (LTA). Our findings demonstrated that >52% maximal platelet aggregation using Plateletworks® had a sensitivity, specificity, and likelihood ratio of 80%, 70%, and 2.67, respectively, in predicting ASA non-sensitivity. This COP was validated in a secondary cohort of 40 patients prescribed 81 mg of ASA using Plateletworks® and LTA. Our data demonstrated that our established COP had a 91% sensitivity and 69% specificity in identifying ASA non-sensitivity using Plateletworks®. In summary, Plateletworks® is a point-of-care platelet function test that can appropriately diagnose ASA non-sensitive patients with a sensitivity exceeding 80%.