Merged Long-Term Data Sets from TOMS and SBUV Total Ozone Measurements

2000 
Total ozone has been measured by a series of nadir-viewing satellite instruments. These measurements begin with the Total Ozone Mapping Spectrometer (TOMS) and Solar Backscatter UltraViolet (SBUV) instruments on Nimbus 7, launched in late 1978. The measurements have continued with the Meteor 3 TOMS, Earth Probe TOMS, and NOAA 9,11,14 SBUV/2 instruments. The problem for producing a long-term data set is establishing the relative calibration of the various instruments to better than 1%. There was a nearly two year gap between the Meteor 3 TOMS and the Earth Probe TOMS. This gap is filled by the NOAA 9 and 11 SBUV/2 instruments, but they were in drifting orbits that result in effective gaps in the record when the equator crossing time occurs at large solar zenith angle. We have used recently re-derived calibrations of the SBUV instruments using the D-pair (306/313 nm wavelengths) data at the equator. These equatorial D-pair measurements should maintain the internal calibration of each instrument better than previous approaches. We then use the comparisons between instruments during their overlap periods to establish a consistent calibration over the entire data set. The resulting merged ozone data set is independent of the ground-based Dobson/Brewer network.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []