Effect of Reader and Patient Parameters on the Performance of Last-Image-Hold for Fluoroscopic Grading of Vesicoureteric Reflux

2019 
OBJECTIVE. The purpose of this study is to determine the effect of different reader and patient parameters on the degree of agreement and the rate of misclassification of vesicoureteric reflux grading on last-image-hold frames in relation to spot-exposed frames from voiding cystourethrography (VCUG) as well as to determine the nature of reflux misclassification on last-image-hold frames. MATERIALS AND METHODS. Blinded readers conducted a retrospective evaluation of last-image-hold and spot-exposed frames of the renal fossae from 191 sequential VCUG examinations performed during a five-year period. Kappa tests were used to determine the agreement between reflux gradings and to assess the impact of reader and patient parameters. Pearson product-moment correlations were used to evaluate the effect of patient parameters on reader level of certainty regarding reflux grading. RESULTS. We measured almost perfect overall agreement for more experienced readers and substantial overall agreement for less experienced...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    13
    References
    0
    Citations
    NaN
    KQI
    []