Intrapulmonary Vein Ablation Without Stenosis: A Novel Balloon‐Based Direct Current Electroporation Approach

2018 
Background Current thermal ablation methods for atrial fibrillation, including radiofrequency and cryoablation, have a suboptimal success rate. To avoid pulmonary vein (PV) stenosis, ablation is performed outside of the PV, despite the importance of triggers inside the vein. We previously reported on the acute effects of a novel direct current electroporation approach with a balloon catheter to create lesions inside the PVs in addition to the antrum. In this study, we aimed to determine whether the effects created by this nonthermal ablation method were associated with irreversible lesions and whether PV stenosis or other adverse effects occurred after a survival period. Methods and Results Initial and survival studies were performed in 5 canines. At the initial study, the balloon catheter was inflated to contact the antrum and interior of the PV. Direct current energy was delivered between 2 electrodes on the catheter in ECG‐gated 100 μs pulses. A total of 10 PVs were treated demonstrating significant ac...