eBP: FREQUENT AND COMFORTABLE BLOOD PRESSURE MONITORING FROM INSIDE HUMAN'S EARS

2020 
Diagnosing hypertension or hemodialysis requires patients to carry a blood pressure (BP) monitoring device for 24 hours. Th erefore, wearing the wrist/arm-based BP monitoring device, in this case, has a signifi cant impact on users' daily activities. To address the problem, we developed eBP, an ear-worn device that measures blood pressure from inside the ear. Th rough the evaluation of 35 subjects, eBP can achieve the average error of 1.8 mmHg for systolic BP and -3.1 mmHg for diastolic BP with the standard deviation error of 7.2 mmHg and 7.9 mmHg, respectively.