Detection of malaria parasites in samples from returning US travelers using the Alethia® Malaria Plus LAMP assay.

2021 
Objective In this study, the performance of a commercially available malaria LAMP assay (Alethia® Malaria Plus LAMP) was evaluated using retrospective clinical samples obtained from travelers returning to the United States of America (USA). Recently, several laboratories in non-malaria endemic countries evaluated the use of the loop mediated isothermal amplification (LAMP) assays for the diagnosis of imported malaria cases. These tests are simpler than polymerase-chain reaction (PCR)-based assays and were shown to have high sensitivity. Much of malaria diagnoses in the USA, is undertaken at the state level using mainly microscopy and rapid diagnostic tests (RDTs). However, molecular tools offer greater sensitivity over microscopy and RDTs. A reliable, easy to perform molecular assay can provide a test of choice for the accurate detection of malaria parasites in places where expert microscopy is lacking and/or for the detection of low-parasite density infections. Results The Alethia® Malaria Plus LAMP assay was easy to use, had similar test performances as the real-time PCR reference test and results were obtained faster (within 1 h) than the reference test. The sensitivity of the assay was 100% with a kappa score of 1 when compared to the reference PET-PCR assay.