Precision of computer-aided volumetry of artificial small solid pulmonary nodules in ex vivo porcine lungs

2007 
The purpose of this study was to investigate the precision of CT-based volumetric measurements of artificial small pulmonary nodules under ex vivo conditions. We implanted 322 artificial nodules in 23 inflated ex vivo porcine lungs in a dedicated chest phantom. The lungs were examined with a multislice spiral CT (20 mAs, collimation 16×0.75 mm, 1 mm slice thickness, 0.7 mm increment). A commercial volumetry software package (LungCARE VA70C-W; Siemens, Erlangen, Germany) was used for volume analysis in a semi-automatic and a manual corrected mode. After imaging, the lungs were dissected to harvest the nodules for gold standard determination. The volumes of 202 solitary, solid and well-defined lesions without contact with the pleura, greater bronchi or vessels were compared with the results of volumetry. A mean nodule diameter of 8.3 mm (±2.1 mm) was achieved. The mean relative deviation from the true lesion volume was –9.2% (±10.6%) for semi-automatic and –0.3% (±6.5%) for manual corrected volumetry. The s...