Usefulness of Attenuation Imaging with an Ultrasound Scanner for the Evaluation of Hepatic Steatosis
2019
Abstract We investigated the diagnostic capability of the proprietary attenuation imaging (ATI) modality found on some Canon Medical Systems Corp. ultrasound scanners to detect histologically diagnosed steatosis in 148 patients. ATI values increased significantly with increasing steatosis grade ( p 66% of hepatocytes) were 0.85, 0.91 and 0.91. In addition, ATI values increased significantly with increasing steatosis grades ( p = 0.002) even in obese patients (n = 41). The diagnostic values of ATI for steatosis grades ≥ 1, ≥ 2 and 3 in obese patients were 0.72, 0.72 and 0.78. Furthermore, ATI values increased significantly with increasing steatosis grade ( p