Identification of malignant early repolarization pattern by late QRS activity in high‐resolution magnetocardiography
2020
BACKGROUND: The early repolarization pattern (ERP) in electrocardiography (ECG) has been considered as a risk for ventricular fibrillation (VF), but effective methods for identification of malignant ERP are still required. We investigated whether high spatiotemporal resolution 64-channel magnetocardiography (MCG) would enable distinction between benign and malignant ERPs. METHODS: Among all 2,636 subjects who received MCG in our facility, we identified 116 subjects (43 +/- 18 years old, 54% male) with inferior and/or lateral ERP in ECG and without structural heart disease, including 13 survivors of VF (ERP-VF(+)) and 103 with no history of VF (ERP-VF(-)). We measured the following MCG parameters in a time-domain waveform of relative current magnitude: (a) QRS duration (MCG-QRSD), (b) root-mean-square of the last 40 ms (MCG-RMS40), and (c) low amplitude ( /= 100 ms and MCG-RMS40 = 0.24 were 69%, 74%, and 6.33 (95% CI, 1.80-22.3), and 92%, 48%, and 10.9 (95% CI, 1.37-86.8), respectively. CONCLUSION: Magnetocardiography is an effective tool to distinguish malignant and benign ERPs.