A standard operating procedure for assessing liquid handler performance in high-throughput screening

2002 
The thrust of early drug discovery in recent years has been toward the configuration of homogeneous miniaturized assays. This has allowed organizations to contain costs in the face of exponential increases in the number of screening assays that need to be run to remain competitive. Miniaturization brings with it an increasing dependence on instrumentation, which over the past several years has seen the development of nanodispensing capability and sophisticated detection strategies. To maintain confidence in the data generated from miniaturized assays, it is critical to ensure that both compounds and reagents have been delivered as expected to the target wells. The authors have developed a standard operating procedure for liquid-handling quality control that has enabled them to evaluate performance on 2 levels. The first level provides for routine daily testing on existing instrumentation, and the second allows for more rigorous testing of new dispensing technologies. The procedure has shown itself to be u...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    37
    References
    47
    Citations
    NaN
    KQI
    []