In vitro feasibility of a sac-sealing endoprosthesis in a double chimney graft configuration for juxtarenal aneurysm.

2014 
Purpose:To investigate in an in vitro juxtarenal aneurysm flow model the feasibility and efficacy of using a sac-sealing endoprosthesis in a chimney graft configuration.Methods:In two experiments, a Nellix sac-sealing endoprosthesis was used as the main graft in a double-branched chimney graft configuration, using a self-expanding Viabahn stent-graft and a balloon-expandable Advanta V12 stent-graft in a pressurized silicone juxtarenal aneurysm flow model. In two consecutive experiments, the chimney graft balloons were inflated (1) at the beginning of and (2) to simulate varying renal ischemic times half-way through the injection of the sac-sealing polymer into the Nellix endobags. The balloons and were kept inflated until the endobags were filled and the polymer was cured. The aneurysm model was connected to a roller pump, pumping gelatin-water at a rate of 100 beats per minute. Before and after 24 hours of continuous flow, computed tomography (CT) scans were made using contrast injection. The CT scans we...