Clinical Application of Non-Contrast-Enhanced Dixon Water-Fat Separation Compressed SENSE Whole-Heart Coronary MR Angiography at 3.0 T With and Without Nitroglycerin.

2021 
BACKGROUND 3.0 T non-contrast-enhanced nitroglycerin (NTG)-assisted whole-heart coronary magnetic resonance angiography (MRA) employing Dixon water-fat separation and compressed SENSE (CS-SENSE) acceleration is a promising method for diagnosing coronary artery disease (CAD). PURPOSE To evaluate the diagnostic performance of this technique for detecting clinically-relevant (≥50% diameter reducing) CAD and to evaluate the difference in NTG-induced coronary vasodilation between patients with and without clinically-relevant CAD. STUDY TYPE Prospective. POPULATION Sixty-six patients with suspected CAD. FIELD STRENGTH/SEQUENCE 3.0 T; CSSENSE, Dixon water-fat separation, three-dimensional segmented turbo field gradient-echo sequence for whole-heart coronary MRA. ASSESSMENT Overall image quality of coronary MRA was calculated on the basis of all visible coronary segments. The diagnostic performance of coronary MRA for detecting a ≥50% reduction in coronary artery diameter with and without NTG was compared using X-ray coronary angiography (CAG) as the reference. According to CAG, patients were divided into a non-clinically-relevant CAD group and clinically-relevant CAD group, and the difference in NTG-induced vasodilation between the groups was evaluated. STATISTICAL TESTS Unpaired/paired Student's t-test, Mann-Whitney U test, paired Wilcoxon signed-rank test, χ2 test, McNemar test. A two-tailed P value <0.05 was considered significant. RESULTS Overall image quality was increased significantly in the coronary MRA images after NTG. The diagnostic performance of the non-NTG vs. NTG-assisted coronary MRA was as follows on a per-patient basis: sensitivity 94.3% vs. 94.3%, specificity 64.5% vs. 83.9%, positive predictive value 75.0% vs. 86.8%, negative predictive value 90.9% vs. 92.9%, and accuracy 80.3% vs. 89.4%, respectively. NTG-induced vasodilation was significantly lower in the clinically-relevant CAD group than in the non-clinically-relevant CAD group (13.7 ± 8.1% vs. 24.1 ± 16.3%). DATA CONCLUSION Non-contrast Dixon water-fat separation CS-SENSE coronary MRA at 3.0 T can noninvasively detect clinically-relevant CAD and sublingual NTG improved performance. Combining pre- and post-NTG coronary MRA may provide a simple noninvasive and nonionizing test to evaluate coronary vasodilation function. LEVEL OF EVIDENCE 1 TECHNICAL EFFICACY STAGE: 2.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    37
    References
    1
    Citations
    NaN
    KQI
    []