Methods and potentials of magnetic resonance imaging for monitoring radiofrequency hyperthermia in a hybrid system.
2005
Introduction: Non-invasive thermometry (NIT) is a valuable and probably indispensable tool for further development of radiofrequency (RF) hyperthermia. A hybridization of an MRI scanner with a hyperthermia system is necessary for a real-time NIT. The selection of the best thermographic method is difficult, because many parameters and attributes have to be considered.Methods: In the hybrid system (Siemens Symphony/BSD-2000-3D) the standard methods for NIT were tested such as T1, diffusion (ADC: apparent diffusion coefficient) and proton-resonance-frequency shift (PFS) method. A series of three-dimensional datasets was acquired with different gradient-echo sequences, diffusion-weighted EPI spin-echo sequences and calculated MR-temperatures in the software platform AMIRA-HyperPlan. In particular for the PFS-method, corrective methods were developed and tested with respect to drift and other disturbances. Experiments were performed in phantoms and the results compared with direct temperature measurements. The...