Wear of different materials for total hip replacement under adverse stop-dwell-start in vitro wear simulation conditions:

2018 
Hip simulation is a common technique for pre-clinical evaluation of wear performance of total hip arthroplasty. Standard techniques replicate kinematics of walking patterns of a typical patient. Attention has focussed in developing simulations of other typical patient daily activities to improve accuracy of wear predictions. A method for simulating stop-dwell-start motion during patient walking and the effect on 36-mm metal-on-metal total hip arthroplasty was previously presented by the authors. This study sought to extend the previous work to look at the effect of these conditions on ceramic-on-ceramic, metal-on-polyethylene and ceramic-on-polyethylene bearings. Two stop-dwell-start protocols were used: one reproducing average patient movement patterns and one examining more severe conditions. For all materials tested, no significant increase in wear was observed under average stop-dwell-start conditions, suggesting the bearing types tested are robust to this type of activity. A significant increase in w...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    21
    References
    8
    Citations
    NaN