Patient dose in digital mammography.

2004 
In the present investigation, we analyze the dose of 5034 patients (20 137 images) who underwent mammographic examinations with a full-field digital mammography system. Also, we evaluate the system calibration by analyzing the exposure factors as a function of breast thickness. The information relevant to this study has been extracted from the image DICOM header and stored in a database during a 3-year period (March 2001–October 2003). Patient data included age, breast thickness, kVp, mAs, target/filter combination, and nominal dose values. Entrance surface air kerma (ESAK) without backscatter was calculated from the tube output as measured for each voltage used under clinical conditions and from the tube loading (mAs) included in the DICOM header. Mean values for the patient age and compressed breast thickness were 56 years (SD: 11) and 52 mm (SD: 13), respectively. The majority of the images was acquired using the STD (for standard) automatic mode (98%). The most frequent target/filter combination automatically selected for breast smaller than 35 mm was Mo/Mo (75%); for intermediate thicknesses between 35 and 65 mm, the combinations were Mo/Rh (54%) and Rh/Rh (38.5%); Rh/Rh was the combination selected for 91% of the cases for breasts thicker than 65 mm. A wide kVp range was observed for each target/filter combination. The most frequent values were 28 kVp for Mo/Mo, 29 kVp for Mo/Rh, and 29 and 30 kV for Rh/Rh. Exposure times ranged from 0.2 to 4.2 s with a mean value of 1.1 s. Average glandular doses (AGD) per exposure were calculated by multiplying the ESAK values by the conversion factors tabulated by Dance for women in the age groups 50 to 64 and 40 to 49. This approach is based on the dependence of breast glandularity on breast thickness and age. The total mean average glandular dose ( AGD T ) was calculated by summing the values associated with the pre-exposure and with the main exposure. Mean AGD T per exposure was 1.88 mGy (CI 0.01) and the mean AGD T per examination was 3.8 mGy, with 4 images per examination on average. The mean dose for cranio-caudal view (CC) images was 1.8 mGy, which is lower than that for medio-lateral oblique (MLO) view because the thickness for CC images was on average 10% lower than that for MLO images. Mean AGD T for the oldest group of women (1.90) was 3% higher than the AGD T for the younger group (1.85) due to the larger compressed breast thickness of women in the older group (10% on average). Differences between the corresponding AGD T values of each age group were lowest for breast thicknesses in the range 40–60 mm, being slightly higher for the women in the older group.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    18
    References
    50
    Citations
    NaN
    KQI
    []