Multicenter evaluation of the Elecsys Toxo IgG and IgM tests for the diagnosis of infection with Toxoplasma gondii.

2015 
Detection of IgG and IgM antibodies is commonly performed for the diagnosis of infection with Toxoplasma gondii. We determined the accuracy of the Elecsys Toxo IgG and IgM test at four European laboratories compared to local reference methods. Coefficients of variation for reproducibility ranged from 1.0 to 6.5% for IgG and from 0.8 to 3.2% for IgM. Seroconversion panels revealed high overall concordance with the reference tests. The Elecsys test detected IgG antibodies earlier than the Cobas Core IgG test in 19 of 47 panels; persisting IgM antibodies were observed in the VIDAS but not the Elecsys test in five of 47 panels. In 31.4% of latent stage sera with persistent IgM antibodies (positive LIASON IgM), the Elecsys IgM test gave negative results indicating increased “clinical” specificity. Sensitivity and specificity of the Elecsys IgG assay ranged from 99.45 to 100% and 87.50–99.80%, respectively, and 91.11–95.74 and 98.45–99.79% for the Elecsys IgM assay, respectively. In conclusion, excellent reproducibility and accuracy make the Elecsys Toxo G and M tests highly suitable for the detection of anti-T. gondii IgG and IgM antibodies. The lower detection rates for persistent IgM in the Elecsys IgM test increase “clinical” specificity and decrease the need for follow-up testing.