Comparison of Two Commercial Matrix-Assisted Laser Desorption/Ionization-Time of Flight Mass Spectrometry (MALDI-TOF MS) Systems for Identification of Nontuberculous Mycobacteria

2019 
OBJECTIVES: This multicenter study's aim was to assess the performance of two commercially available matrix-assisted laser desorption/ionization time of flight mass spectrometry systems in identifying a challenge collection of clinically relevant nontuberculous mycobacteria (NTM). METHODS: NTM clinical isolates (n = 244) belonging to 23 species/subspecies were identified by gene sequencing and analyzed using Bruker Biotyper with Mycobacterial Library v5.0.0 and bioMerieux VITEK MS with v3.0 database. RESULTS: Using the Bruker or bioMerieux systems, 92% and 95% of NTM strains, respectively, were identified at least to the complex/group level; 62% and 57%, respectively, were identified to the highest taxonomic level. Differentiation between members of Mycobacterium abscessus, M fortuitum, M mucogenicum, M avium, and M terrae complexes/groups was problematic for both systems, as was identification of M chelonae for the Bruker system. CONCLUSIONS: Both systems identified most NTM isolates to the group/complex level, and many to the highest taxonomic level. Performance was comparable.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    46
    References
    12
    Citations
    NaN
    KQI
    []