Reverse Engineering Nature to Design Anatomic Articular Surfaces for Total Knee Arthroplasty Implants

2018 
Introduction:While kinematic abnormalities of contemporary TKA implants have been well established, a solution has not yet been achieved. We hypothesized that contemporary TKA implants are not compatible with normal soft-tissue function and normal knee motion. We propose a novel technique for reverse engineering advanced implant articular surfaces (biomimetic surface), by using accurate 3D kinematics of normal knees. This technique accounts for surgical placement of the implants, and allows design of tibial and femoral articular surfaces in conjunction.Methods:Magnetic resonance imaging was used to create 3D knee models of 40 normal subjects (24 male, 16 female, age 29.9 ± 9.7 years), and bi-planar fluoroscopy was used to capture 3D knee motion during a deep knee bend. These data were combined to create a 3D virtual representation of an average normal knee and its motion pathway. A TKA femoral component was mounted on the average knee, and moved through its normal kinematic pathway to carve out an articul...