Evaluation of a human glycated hemoglobin test in canine diabetes mellitus

2019 
Glycated hemoglobin A1c (HbA1c) is widely used for monitoring and diagnosing human diabetes mellitus, but is rarely used in veterinary clinics. The goal of our study was to validate the commercial HbA1c testing system SD A1cCare analyzer (Bionote, Gyeoggi-do, South Korea) for use in dogs. Dogs were recruited with owner’s consent. Diabetic status was determined based on clinical signs, fasting hyperglycemia, and glycosuria. Intra-assay precision and linearity were evaluated with EDTA, heparin, or citrate as anticoagulants, and had excellent precision with mean coefficients of variation (CVs) of 2.47%, 2.26%, and 1.92%, respectively. Diluted anticoagulated blood samples showed excellent linear relationships with R2 of 0.991, 0.996, and 0.994, respectively. Inter-assay precision revealed that the mean CV of the normal control was 2.18% and that of the high control was 2.01% (30 repeats). Observed total error of a normal control was 7.81%, and 6.12% for the high control. HbA1c level measured before and after ...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    40
    References
    2
    Citations
    NaN
    KQI
    []