LC-FAIMS-MS/MS for Quantification of a Peptide in Plasma and Evaluation of FAIMS Global Selectivity from Plasma Components

2008 
As a continuation of the evaluation of the utility of high-field asymmetric waveform ion mobility spectrometry (FAIMS) in quantitative bioanalysis, we have developed a sensitive and selective method for the quantification of a peptide drug candidate in rat plasma using FAIMS coupled with liquid chromatography tandem mass spectrometry (LC−MS/MS). The LC-FAIMS-MS/MS method provided significant advantage over the corresponding LC-MS/MS method by reducing chemical/endogenous background noise associated with plasma matrix, thereby improving the sensitivity via increasing the signal-to-noise ratio. Linearity was established within 1−1000 nM in rat plasma, and the overall method accuracy and precision were good meeting the generally adopted acceptance criteria for a bioanalytical method. In a related investigation, we demonstrated the global selectivity of FAIMS from plasma endogenous components as a function of the compensation voltage (CV) across molecular masses that encompass small-molecule drugs. This work ...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    19
    References
    66
    Citations
    NaN
    KQI
    []