Localization of the inferior alveolar canal using ultralow dose CT with iterative reconstruction techniques

2018 
Objectives:To compare subjective and objective localization of the inferior alveolar canal (IAC) on multidetector CT (MDCT) images obtained by ultralow doses in combination with the reconstruction techniques of filtered backprojection (FBP), adaptive statistical iterative reconstruction (ASIR), or model-based iterative reconstruction (MBIR) as compared to standard dose MDCT and FBP.Methods:Three cadavers were imaged with a reference standard dose MDCT examination (volume CT dose index: 29.4 mGy) reconstructed with FBP and 5 low dose protocols (LD1-5) (volumeCT dose index: 4.19, 2.64, 0.99, 0.53, 0.29 mGy) reconstructed with FBP, ASIR 50, ASIR 100, and MBIR. Linear measurements from the crest of the ridge to the roof of the IAC were recorded. The results from the test protocols were compared with those from the reference using Bland–Altman plots.Results:Only three test protocols allowed the identification of the position of the IAC on all the sample sites: LD1/FBP and LD1/ASIR 100 and LD2/FBP. All three pr...