iStent versus iStent inject implantation combined with phacoemulsification in open angle glaucoma.

2021 
Purpose To compare the outcomes of iStent vs. iStent inject implantation combined with phacoemulsification. Methods This single center retrospective comparative case series included subjects with open angle glaucoma who underwent iStent or iStent inject implantation combined with phacoemulsification with ≥1 year follow-up. The main outcome measures were in-group and between-group changes in intraocular pressure (IOP) and medication number, proportion of eyes that achieved IOP ≤15 mmHg, and surgical success defined as 20% IOP reduction from baseline at 6/12 months. Univariate/multivariate regression analyses were done to identify predictors of surgical failure. Results One hundred ninety-seven eyes of 148 patients were included (122 iStent, 75 iStent inject). Both groups achieved significant IOP and medication reduction at months 6/12 (P 0.05). More iStent inject eyes achieved IOP ≤15 mmHg at month 6 (P = 0.003) and 12 (P = 0.047). Surgical success was comparable in both groups at months 6/12 (P > 0.05). Kaplan-Meier survival analysis showed similar cumulative rate of surgical failure at year-1 in both groups (P = 0.644). The multivariate model identified older age (P = 0.017) and lower baseline IOP (P = 0.002) as the strongest predictors of surgical failure. Conclusion Compared to iStent, iStent inject achieved lower IOP at month 6 and higher proportion of eyes achieved IOP ≤15 mmHg at month 6/12. However, surgical success was similar in both groups. Predictors of surgical failure were older age and lower baseline IOP rather than the stent type.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []