Preoperative staging of rectal cancer using a 7.5 MHz front-loading US probe

2000 
Background: Conventional echoendoscopes have disadvantages when used for staging colorectal cancer including the inability to pass the instrument through tight stenosis and limited maneuverability. This study evaluated the preoperative use of a newly developed 7.5 MHz front-loading ultrasound probe (FLUP) for local staging of rectal cancer. Methods: A 7.5 MHz FLUP, diameter 7.3 mm, was used in this study. The mechanical shaft portion of the probe can be passed in retrograde fashion through the accessory channel of a standard colonoscope. Thirty-nine patients with rectal cancer underwent ultrasonography with this probe. The tumors were staged using the TNM system, and the results were compared with the histologic findings of the resected specimens. Results: The FLUP proved to be satisfactory, with respect to maneuverability, for traversing stenosis and accurate recognition of small tumors under direct endoscopic control. The accuracy of the FLUP for T staging was 82% (32 of 39) for all tumors, 90% in pT1, and 79% in pT2 to pT4 tumors. The accuracy of the FLUP for N staging was 72% (23 of 32) overall. The sensitivity was 83%, the specificity was 65%, the positive predictive value was 59%, and the negative predictive value was 87%. Conclusions: The 7.5 MHz FLUP appears to be useful for preoperative local staging of rectal cancer. This system makes it technically easier to image small cancers as well as advanced rectal cancers.