Early Vascular Healing Following Bioresorbable-Polymer Sirolimus-Eluting Stent Placement Compared to That with Durable-Polymer Everolimus-Eluting Stent: Sequential Optical Coherence Tomography Study

2021 
A recent thinner strut drug-eluting stent might facilitate early strut coverage after its placement. We aimed to investigate early vascular healing responses after the placement of an ultrathin-strut bioresorbable-polymer sirolimus-eluting stent (BP-SES) compared to those with a durable-polymer everolimus-eluting stent (DP-EES) using optical coherence tomography (OCT) imaging.This study included 40 patients with chronic coronary syndrome (CCS) who underwent OCT-guided percutaneous coronary intervention (PCI). Twenty patients each received either BP-SES or DP-EES implantation. OCT was performed immediately after stent placement (baseline) and at 1-month follow-up.At one month, the percentage of uncovered struts reduced significantly in both the BP-SES (80.9 ± 10.3% to 2.9 ± 1.7%; P < 0.001) and DP-EES (81.9 ± 13.0% to 5.7 ± 1.8%; P < 0.001) groups, and the percentage was lower in the BP-SES group than in the DP-EES group (P < 0.001). In the BP-SES group, the percentage of malapposed struts also decreased significantly at 1 month (4.9 ± 3.7% to 2.6 ± 3.0%; P = 0.025), which was comparable to that of the DP-EES group (2.5 ± 2.2%; P = 0.860). The optimal cut-off value of the distance between the strut and vessel surface immediately after the placement to predict resolved malapposed struts was ≤ 160 μm for BP-SES and ≤ 190 μm for DP-EES.Compared to DP-EES, ultrathin-strut BP-SES demonstrated favorable vascular responses at one month, with a lower rate of uncovered struts and a comparable rate of malapposed struts.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []