Multisection FLASH: method for breath-hold MR imaging of the entire liver.

1992 
One hundred ten patients with various focal liver lesions were imaged with a multisection fast low-angle shot (FLASH) gradient-echo sequence with an echo time of 4.6 msec. This sequence enabled the acquisition of 19 T1-weighted magnetic resonance (MR) images of the liver within a single 26-second breath hold. Patients were also examined with standard T1- and T2-weighted spin-echo (SE) sequences. The multisection FLASH sequence provided significantly higher (P less than .01) liver-spleen contrast, liver-spleen signal-difference-to-noise ratio (SD/N), liver-tumor contrast, and liver-tumor SD/N than the T1-weighted SE sequence but lower values than the T2-weighted SE sequence. Motion artifacts were reduced with the multisection FLASH sequence compared with both SE sequences (P less than .01). The overall image quality of the multisection FLASH images was similar to that of the T1-weighted SE images and superior to that of T2-weighted SE images. The most important characteristics of the multisection FLASH tec...