DCO 2 /PaCO 2 correlation on high-frequency oscillatory ventilation combined with volume guarantee using increasing frequencies in an animal model
2019
To examine the correlation DCO2/PaCO2 on high-frequency oscillatory ventilation (HFOV) combined with volume guarantee (VG) throughout increasing frequencies in two different respiratory conditions, physiological and low compliance. Neonatal animal model was used, before and after a bronchoalveolar lavage (BAL). HFOV combined with VG was used. The frequency was increased from 10 to 20 Hz, and high-frequency tidal volume (VThf) was gradually decreased maintaining a constant DCO2. Arterial partial pressure of carbon dioxide (PaCO2) was evaluated after each frequency and VThf change. Six 2-day-old piglets were studied. A linear decrease in PaCO2 was observed throughout increasing frequencies in both respiratory conditions while maintaining a constant DCO2, showing a significant difference between the initial PaCO2 (at 10 Hz) and the PaCO2 obtained at 18 and 20 Hz. A new DCO2 equation (corrected DCO2) was calculated in order to better define the correlation between DCO2 and the observed PaCO2.