Ankle/Brachial Index in the Primary Care Setting

2004 
Peripheral arterial disease (PAD) is an underdiagnosed circulatory problem in the primary care setting. Individuals are at increased risk for cardiovascular disease; therefore, there is the need for a technique capable of early identification and detection of patients with PAD. The focus of this study was to compare the accuracy of a new operator-independent method of measuring ankle brachial index (ABI) with the traditional Doppler ultrasound method of determining ABI. In 246 limbs the authors compared ankle systolic pressures and ABI measured by a new automated oscillatory method called the ABIgram with those measured by standard Doppler ultrasound. In phase 1, the 2 methods for measuring ankle systolic pressure had a mean difference of 2 mm Hg with a standard deviation of 6.7 mm Hg. In phase 2 the mean difference was 3.1 mm Hg with a standard deviation of 5.1 mm Hg. Further, ABI as measured by the 2 methods fell within 1% and demonstrated a 5% error in reproducibility. These numbers pass the SP-10 stan...