Bioprosthetic Valve Fracture: Technical Insights from a Multicenter Study

2019 
Abstract Objective Valve-in-valve transcatheter aortic valve replacement (VIV TAVR) can result in high residual gradients that are associated with increased mortality. Bioprosthetic valve fracture (BVF) has been shown to improve residual gradients following VIV TAVR; however, factors influencing the results of BVF have not been studied. Methods BVF was performed in 75 patients at 21 centers. Hierarchical multiple linear regression was performed to identify variables that were associated with lower final transvalvular gradient. Results Surgical valves with a median true internal diameter of 18.5 mm (interquartile range, 17.0-20.5 mm) were treated with VIV TAVR in conjunction with BVF using balloon-expandable (n = 43) or self-expanding (n = 32) transcatheter heart valves with a median size of 23 mm (interquartile range, 23-23 mm). There were no aortic root disruptions, coronary occlusions, or new pacemakers; in-hospital or 30-day mortality was 2.6% (2 out of 75). Final mean transvalvular gradient was 9.2 ± 6.3 mm Hg, but was significantly lower when BVF was performed after VIV TAVR compared with BVF first (8.1 ± 4.8 mm Hg vs 16.9 ± 10.1 mm Hg; P P P  = .038) were the only independent predictors of lower final mean gradient. Conclusions BVF can be performed safely and results in reduced residual transvalvular gradients. Performing BVF after VIV TAVR and using larger balloon appears to achieve the best hemodynamic results.