Dose reduction in 64-row whole-body CT in multiple trauma: an optimized CT protocol with iterative image reconstruction on a gemstone-based scintillator.

2016 
Objective:Evaluation of potential dose savings by implementing adaptive statistical iterative reconstruction (ASiR) on a gemstone-based scintillator in a clinical 64-row whole-body CT (WBCT) protocol after multiple trauma.Methods:Dose reports of 152 WBCT scans were analysed for two 64-row multidetector CT scanners (Scanners A and B); the main scanning parameters were kept constant. ASiR and a gemstone-based scintillator were used in Scanner B, and the noise index was adjusted (head: 5.2 vs 6.0; thorax/abdomen: 29.0 vs 46.0). The scan length, CT dose index (CTDI) and dose–length product (DLP) were analysed. The estimated mean effective dose was calculated using normalized conversion factors. Student's t-test was used for statistics.Results:Both the mean CTDI (mGy) (Scanner A: 53.8 ± 2.0, 10.3 ± 2.5, 14.4 ± 3.7; Scanner B: 48.7 ± 2.2, 7.1 ± 2.3, 9.1 ± 3.6; p < 0.001, respectively) and the mean DLP (mGy cm) (Scanner A: 1318.9 ± 167.8, 509.3 ± 134.7, 848.8 ± 254.0; Scanner B: 1190.6 ± 172.6, 354.6 ± 128.3, 56...