Effect of Bipolar Electrode Spacing on Phrenic Nerve Stimulation and Left Ventricular Pacing Thresholds

2012 
Background—Phrenic nerve stimulation (PNS) is a common complication of cardiac resynchronization therapy when left ventricular (LV) pacing occurs via a coronary vein. The purpose of this study was to evaluate the effects of bipolar electrode spacing on PNS and LV pacing thresholds. Methods and Results—Electrophysiology catheters with standard (2 mm-5 mm-2 mm) or modified (1 mm-5 mm-1 mm) interelectrode spacing was, respectively, inserted in a posterior/lateral cardiac vein in a randomized order in 6 anesthetized dogs via jugular access. The phrenic nerve was dissected via a left minithoracotomy and repositioned over the vein as close as possible to one of the electrodes. The presence of PNS was verified (ie, PNS threshold <2 V at 0.5 ms in unipolar configuration). Bipolar pacing was delivered using the electrode closest to the phrenic nerve as the cathode, and multiple bipolar electrode spacing configurations were tested. During bipolar pacing, PNS threshold increased as bipolar electrode spacing was redu...