Effects of tube voltage and iodine contrast medium on radiation dose of whole-body CT.

2021 
BACKGROUND The low-tube-voltage scan generally needs a higher tube current than the conventional 120 kVp to maintain the image noise. In addition, the low-tube-voltage scan increases the photoelectric effect, which increases the radiation absorption in organs. PURPOSE To compare the organ radiation dose caused by iodine contrast medium between low tube voltage with low contrast medium and that of conventional 120-kVp protocol with standard contrast medium. MATERIAL AND METHODS After the propensity-matching analysis, 66 patients were enrolled including 33 patients with 120 kVp and 600 mgI/kg and 33 patients with 80 kVp and 300 mgI/kg (50% iodine reduction). The pre- and post-contrast phases were assessed in all patients. The Monte Carlo simulation tool was used to simulate the radiation dose. The computed tomography (CT) numbers for 10 organs and the organ doses were measured. The organ doses were normalized by the volume CT dose index, and the 120-kVp protocol was compared with the 80-kVp protocol. RESULTS On contrast-enhanced CT, there were no significant differences in the mean CT numbers of the organs between 80-kVp and 120-kVp protocols except for the pancreas, kidneys, and small intestine. The normalized organ doses at 80 kVp were significantly lower than those of 120 kVp in all organs (e.g. liver, 1.6 vs. 1.9; pancreas, 1.5 vs. 1.8; spleen, 1.7 vs. 2.0) on contrast-enhanced CT. CONCLUSION The low tube voltage with low-contrast-medium protocol significantly reduces organ doses at the same volume CT dose index setting compared with conventional 120-kVp protocol with standard contrast medium on contrast-enhanced CT.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    29
    References
    0
    Citations
    NaN
    KQI
    []