Effect of rotational prosthetic alignment variation on tibiofemoral contact pressure distribution and joint kinematics in total knee replacement

2017 
In total knee replacement surgery, implant alignment is one of the most important criteria for successful long-term clinical outcome. During total knee replacement implantation, femoral and tibial alignment are determined through appropriate bone resections, which could vary based on patient anatomy, implant design and surgical technique and further influence loading conditions and clinical outcomes. The current research focused on three critical alignment parameters for total knee replacement insertion: femoral component internal/external (I/E) rotation, varus–valgus tibiofemoral angulation and posterior tibial slope. A computational finite element model of total knee replacement implant was developed and validated comparing with kinematic outputs generated from experimentally simulated knee joint motion. The FE model was then used to assess 12 different alignment scenarios based on previous case reports. Postoperative knee kinematics and joint contact pressure during simulated gait motion were assessed....