Abstract Meniscal bearing fracture is a rare complication of phase III Oxford unicompartmental knee replacement (UKR). We report a case of a meniscal bearing fracture that occurred 7 years after phase III Oxford medial UKR. The meniscal bearing showed uneven delamination of the polyethylene in the thinnest articular surface and an impingement lesion. This lesion initiated a fatigue crack that propagated to cause failure of the meniscal bearing. This is the first report of a meniscal bearing fracture without a posterior marker wire.