Transfemoral Transcatheter Tricuspid Valve Replacement With the EVOQUE System: A Multicenter, Observational, First-in-Human Experience.

2021 
Abstract Objectives The purpose of this observational first-in-human experience was to investigate the feasibility and safety of the EVOQUE tricuspid valve replacement system and its impact on short-term clinical outcomes. Background Transcatheter tricuspid intervention is a promising option for selected patients with severe tricuspid regurgitation (TR). Although transcatheter leaflet repair is an option for some, transcatheter tricuspid valve replacement (TTVR) may be applicable to a broader population. Methods Twenty-five patients with severe TR underwent EVOQUE TTVR in a compassionate-use experience. The primary outcome was technical success, with NYHA (NYHA) functional class, TR grade, and major adverse cardiac and cerebrovascular events assessed at 30-day follow-up. Results All patients (mean age 76 ± 3 years, 88% women) were at high surgical risk (mean Society of Thoracic Surgeons risk score 9.1 ± 2.3%), with 96% in NYHA functional class III or IV. TR etiology was predominantly functional, with mean tricuspid annular diameter of 44.8 ± 7.8 mm and mean tricuspid annular plane systolic excursion of 16 ± 2 mm. Technical success was 92%, with no intraprocedural mortality or conversion to surgery. At 30-day follow-up, mortality was 0%, 76% of patients were in NYHA functional class I or II, and TR grade was ≤2+ in 96%. Major bleeding occurred in 3 patients (12%), 2 patients (8%) required pacemaker implantation, and 1 patient (4%) required dialysis. Conclusions This first-in-human experience evaluating EVOQUE TTVR demonstrated high technical success, acceptable safety, and significant clinical improvement. Larger prospective studies are needed to confirm durability and safety and the impact on long-term clinical outcomes.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    27
    References
    25
    Citations
    NaN
    KQI
    []